INVITRO DOPPLER ECHOCARDIOGRAPHIC DETERMINATION OF THE EFFECTIVE ORIFICE AREA OF MECHANICAL HEART-VALVE PROSTHESES

被引:0
|
作者
FEHSKE, W [1 ]
KOHLER, J [1 ]
RABAHIEH, R [1 ]
BRAUN, P [1 ]
HOSTERT, A [1 ]
LUDERITZ, B [1 ]
机构
[1] RHEIN WESTFAL TH AACHEN,INST AERODYNAM,MED STROMUNGSTECH LAB,W-5100 AACHEN,GERMANY
来源
ZEITSCHRIFT FUR KARDIOLOGIE | 1991年 / 80卷 / 07期
关键词
HEART-VALVE PROSTHESES; DOPPLER ECHOCARDIOGRAPHY; EFFECTIVE ORIFICE AREA; CONTINUITY EQUATION; INVITRO MODEL;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In a flow model the effective orifice areas (Ae) of 17 mechanical heart valve prostheses were determined. We measured the Ae-values of several sizes of three types of mechanical prostheses (Medtronic-Hall, St. Jude Medical, and Omnicarbon) under quasi-steady flow conditions using the continuity equation: Ae = flow/maximal transprosthetic velocity. The flow through the model could be determined exactly by directly measuring the decreasing fluid level within the feed tank, while the maximal velocities were calculated from CW-Doppler echocardiographic spectra. It was found that 1) over a range of 200-800 cm3/s Ae was constant for all prostheses and 2) in small aortic prostheses the Ae could be determined with only little scattering of the obtained values, while in large mitral prostheses there was a considerable variation within the results of repeated investigations. For example, in the 21- and 31-Omnicarbon-valves mean values of Ae were calculated as 1.41 and 4.03 cm2, respectively, with standard deviations of 0.05 and 0.49 cm2 as a result of about 70 single calculations in each valve. 3) The absolute values of Ae were smaller than those of comparable in vitro studies based on the Gorlin formula. We conclude that the effective orifice areas of prosthetic heart valves can be easily determined in a flow model by the combination of flow and Doppler echocardiographic measurements. As determinations are based on the same principle, the obtained values should clinically be referred to patients where the corresponding continuity equation for pulsatile flow is used as Ae = stroke volume/time integral of the maximal transvalvular velocity.
引用
收藏
页码:441 / 448
页数:8
相关论文
共 50 条
  • [1] DOPPLER-ECHOCARDIOGRAPHIC RECOGNITION OF NORMAL REGURGITATION IN MECHANICAL HEART-VALVE PROSTHESES
    FEHSKE, W
    KOHLER, J
    KESSEL, D
    RABAHIEH, R
    HAGENDORFF, A
    STEVENS, K
    NIEDEGGEN, A
    LUDERITZ, B
    [J]. ZEITSCHRIFT FUR KARDIOLOGIE, 1992, 81 (11): : 627 - 635
  • [2] DOPPLER-ECHOCARDIOGRAPHIC DETERMINATION OF GEOMETRIC ORIFICE AREAS IN MECHANICAL AORTIC-VALVE PROSTHESES
    HENNEKE, KH
    PONGRATZ, G
    POHLMANN, M
    BACHMANN, K
    [J]. CARDIOLOGY, 1995, 86 (06) : 508 - 513
  • [3] INVITRO HEMOLYSIS BY MECHANICAL HEART-VALVE PROSTHESES WITH TILTING DISK
    WENDT, MO
    POHL, M
    PRATSCH, S
    LERCHE, D
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1992, 15 (11): : 681 - 685
  • [4] VALIDATION OF THE ORIFICE FORMULA FOR ESTIMATING EFFECTIVE HEART-VALVE OPENING AREA
    COCHRANE, T
    KENYON, CJ
    LAWFORD, PV
    BLACK, MM
    CHAMBERS, JB
    SPRIGINGS, DC
    [J]. CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1991, 12 (01): : 21 - 37
  • [5] Effective Orifice Area during Exercise in Bileaflet Mechanical Valve Prostheses
    Bertrand, Philippe B.
    Pettinari, Matteo
    De Canniere, Helene
    Gutermann, Herbert
    Smeets, Christophe J. P.
    Verhaert, David
    Dion, Robert A.
    Verdonck, Pascal
    Vandervoort, Pieter M.
    [J]. JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2017, 30 (04) : 404 - 413
  • [6] HEART-VALVE PROSTHESES AND EXTERNAL MECHANICAL VIBRATIONS
    LAMBERTZ, H
    [J]. DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1990, 115 (07) : 274 - 275
  • [7] CAVITATION DYNAMICS OF MECHANICAL HEART-VALVE PROSTHESES
    LEE, CS
    CHANDRAN, KB
    CHEN, LD
    [J]. ARTIFICIAL ORGANS, 1994, 18 (10) : 758 - 767
  • [8] INVITRO COMPARISON OF AORTIC HEART-VALVE PROSTHESES .1. MECHANICAL VALVES
    KNOTT, E
    REUL, H
    KNOCH, M
    STEINSEIFER, U
    RAU, G
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1988, 96 (06): : 952 - 961
  • [9] INVITRO PULSATILE FLOW HEMODYNAMICS OF 5 MECHANICAL AORTIC HEART-VALVE PROSTHESES
    WALKER, PG
    YOGANATHAN, AP
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1992, 6 : S113 - S123
  • [10] CAVITATION POTENTIAL OF MECHANICAL HEART-VALVE PROSTHESES
    GRAF, T
    FISCHER, H
    REUL, H
    RAU, G
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1991, 14 (03): : 169 - 174