CHANGES IN OXYGEN MEASUREMENTS WHEN WHOLE-BLOOD IS STORED IN ICED PLASTIC OR GLASS SYRINGES

被引:0
|
作者
MAHONEY, JJ [1 ]
HARVEY, JA [1 ]
WONG, RJ [1 ]
VANKESSEL, AL [1 ]
机构
[1] STANFORD UNIV,MED CTR,SCH MED,DIV PULM & CRIT CARE MED,ROOM H3143,STANFORD,CA 94305
关键词
BLOOD GASES; ANALYTICAL ERROR; SAMPLE HANDLING;
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Tonometered whole-blood and plasma specimens were tested in plastic and glass syringes to determine whether clinically significant changes in gas tensions occur during sample storage. When whole blood was tonometered with 60 and 100 mL/L (6% and 10%) oxygen and then stored for 30 min in iced plastic syringes, the p(O2) of the samples remained stable (mean change = +0.4 and +0.8 mmHg, respectively). However, for 140 mL/L (14%) oxygen tonometry, the p(O2) increased significantly (mean change = +8.4 mmHg; P < 0.0001). When tonometered plasma was stored in iced plastic syringes, the p(O2) increased progressively at all three concentrations, with the smallest change occurring at 140 mL/L (mean change = +12.6 mmHg) and the greatest at 60 mL/L oxygen (mean change = +20.9 mmHg). In contrast, when iced glass syringes were used for storing plasma or whole blood, no clinically significant changes in p(O2) were found at any of the tonometered oxygen values for 60 min. When whole blood was stored in plastic syringes at ambient temperature for 30 min, again no clinically significant changes in p(O2) were found at these tonometry conditions. Apparently, some blood gas samples stored in iced plastic syringes may yield clinically significant errors in oxygen tension.
引用
收藏
页码:1244 / 1248
页数:5