The short half-life involves a number of important differences in production, transportation and measurement of radionuclides, and in counting statistics as compared with those in traditional activation analysis. Experiments were performed to investigate the analytical possibilities and prospective utilization of short-lived radionuclides produced by 14-MeV neutron irradiation. A rapid pneumatic transfer system for use with neutron generators was installed and applied for detecting radionuclides with a half-life from 300 ms to 30 s. The transport time for samples with a total mass of 1-4 g is between 130 and 160 ms for pressurized air of 0.1-0.4 MPa. 11 elements were studied by the conventional activation method using both a typical pneumatic transport system (run time 3 s) and the fast pneumatic transport facility. The effect of the cyclic activation technique on the elemental sensitivities was also investigated.