DISTORTIONS IN THE ECG CAUSED BY INSTRUMENTS FOR AMBULATORY ELECTROCARDIOGRAPHY

被引:0
|
作者
BJERREGAARD, P
机构
[1] Univ. Dept. Cardiol., Aarhus Kommunehosp., Aarhus, Denmark
关键词
AMBULATORY MONITORING - ARRHYTHMIA MONITORING - HOLTER TAPES;
D O I
暂无
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The frequency response and speed stability of a widely used tape recorder (Oxford Medilog) and arrhythmia computer (Reynolds Pathfinder) was evaluated and measurements in time and amplitude on the ambulatory ECG compared with correspondent measurements in the conventional ECG. The overall bandwidth was 0. 5-33 Hz (3 dB down) and the mean speed error plus 2% plus or minus 1. 8%. There was no statistically significant difference between P-Q and Q-T intervals measured by the 2 methods, but a significant reduction in wave amplitudes in the ambulatory ECG. The demarcation of QRS complexes was often blurred in the ambulatory ECG and borderline complexes difficult to classify as normal or abnormal on QRS width alone. Artificial J point depression was a common finding in the ambulatory ECG, whereas horizontal or downsloping ST segment depression was faithfully reproduced.