Evaluation of dosimetric performance of "Siemens ONCOR" linear accelerator with small monitor unit values and small segment sizes and consistency with treatment planning system

被引:0
|
作者
Ceylan, Cemile [1 ]
Bas, Hande [1 ]
Kilic, Ayhan [1 ]
Kucuk, Nadir [1 ]
Guden, Metin [1 ]
Engin, Kayihan [1 ]
机构
[1] Anadolu Saglik Merkezi, Radyasyon Onkol Bolumu, Kocaeli, Turkey
来源
关键词
IMRT; small MU; small field size; linearity;
D O I
暂无
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
OBJECTIVES The aim of our study was to evaluate the performance of the "Siemens ONCOR Avant-Garde" linear accelerator under intensity modulated radiation therapy (IMRT) conditions. The beam stability for small monitor units in terms of dose linearity, inter-segment variations and profiles was investigated. Validation of doses at small size segments was done to compare with measurement and calculation dose values. METHODS Measurements of linearity and beam stability with small monitor units were done using 0.6 cc "Farmer" ion chamber and PTW-LA48 linear detector with 47 diodes. Validation of doses for small segment sizes was done for comparison of the values obtained with the PTW Si diode detector and the calculated values on the planning system. RESULTS Dose linearity of small monitor units was found to be <3% for all monitor unit settings. Although inter-segment variations for >2 monitor unit setting were measured at an acceptable value, 3%, <2 monitor unit setting was measured as larger than 3%. Comparison of doses at small segment sizes showed unacceptable value for <2x2 cm(2). Measurements of doses for >2x2 cm(2) were consistent (<3%) with calculated doses from the planning system under the same conditions. CONCLUSION The dose characteristics of Siemens ONCOR Avant-Garde linear accelerator with small monitor units and small segment sizes were found to be suitable for "Step & Shoot" IMRT.
引用
收藏
页码:73 / 79
页数:7
相关论文
共 7 条
  • [1] Dosimetric characteristics of linear accelerator photon beams with small monitor unit settings
    Kang, Sei-Kwon
    Cheong, Kwang-Ho
    Hwang, Taejin
    Cho, Byung Chul
    Kim, Su SSan
    Kim, Kyoung Ju
    Oh, Do Hoon
    Bae, Hoonsik
    Suh, Tae-Suk
    [J]. MEDICAL PHYSICS, 2008, 35 (11) : 5172 - 5178
  • [2] Performance characterization of Siemens Primus linear accelerator under small monitor unit and small segments for the implementation of step-and-shoot intensitymodulated radiotherapy
    Reena, P.
    Dayananda, S.
    Pai, Rajeshri
    Jamema, S. V.
    Gupta, Tejpal
    Deepak, D.
    Rajeev, S.
    [J]. JOURNAL OF MEDICAL PHYSICS, 2006, 31 (04) : 269 - 274
  • [3] Configuration of a treatment planning system: study and dosimetric evaluation for treatments with small fields
    Lobefalo, F.
    Fogliata, A.
    Reggiori, G.
    Stravato, A.
    Cozzi, L.
    Pimpinella, M.
    Palumbo, V.
    Mancosu, P.
    Gaudino, A.
    Scorsetti, M.
    Tomatis, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S413 - S414
  • [4] Dosimetric evaluation of the PEREGRINE IMRT MC treatment planning system at 6 MV for small fields in heterogeneous media
    Sheikh-Bagheri, D
    Cocheci, C
    Heath, E
    Seuntjens, J
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 2016 - 2016
  • [5] A Phantom-Based Analysis of Small Field Monitor Unit Calculation Agreement Between the Eclipse Treatment Planning System and RadCalc Software
    Ellefson, S.
    Clouser, E.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E336 - E336
  • [6] A Comparative Study and Dose Evaluation of Photon Beam for water phantom, 2D-array and Treatment Planning System in Small Field Sizes
    Hassn, Sherihan
    Deiab, Nashaat A.
    Aly, Arafa H.
    [J]. ARAB JOURNAL OF NUCLEAR SCIENCES AND APPLICATIONS, 2020, 53 (01): : 119 - 124
  • [7] The impact of corrected field output factors based on IAEA/AAPM code of practice on small-field dosimetry to the calculated monitor unit in eclipse™ treatment planning system
    Mamesa, Sammuel
    Oonsiri, Sornjarod
    Sanghangthum, Taweap
    Yabsantia, Sumalee
    Suriyapee, Sivalee
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (05): : 65 - 75