Inter-fraction variation in interstitial high-dose-rate brachytherapy

被引:4
|
作者
Kandasamy, Saravanan [1 ]
Reddy, K. [1 ]
Nagarajan, Vivekanandan [2 ]
Vedasoundaram, Parthasarathy [1 ]
Karunanidhi, Gunaseelan [1 ]
机构
[1] JIPMER, Dept Radiotherapy, Reg Canc Ctr, Pondicherry, India
[2] Canc Inst WIA, Med Phys Dept, Madras, Tamil Nadu, India
关键词
high-dose brachytherapy; interfraction in HDR; interstitial implant; positional variation in HDR;
D O I
10.1017/S1460396915000047
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Aim: To evaluate the inter-fraction variation in interstitial high-dose-rate (HDR) brachytherapy. To assess the positional displacement of catheters during the fractions and the resultant impact on dosimetry. Background: Although brachytherapy continues to be a key cornerstone of cancer care, it is clear that treatment innovations are needed to build on this success and ensure that brachytherapy continues to provide quality care for patients. The dosimetric advantages offered by HDR brachytherapy to the tumour volume rely on catheter positions being accurately reproduced for all fractions of treatment. Materials and methods: A total of 66 patients treated over a period of 22 months were considered for this study. All the patients underwent computer tomography (CT) scan and three-dimensional treatment planning was carried out. Brachytherapy treatment was delivered by the HDR afterloading system. On completing the last fraction, CT scan was repeated and treatment re-planning was done. The variation in position of the implanted applicators and their impact on dosimetric parameters were analysed using both the plans. Results: For all breast-implant patients, the catheter displacement and D-90 dose to clinical target volume were <3 mm and 3%, respectively. The displacement for carcinoma of the tongue, carcinoma of the buccal mucosa, carcinoma of the floor of mouth, carcinoma of the cervix, soft-tissue sarcoma and carcinoma of the lip were comparatively high. Conclusion: Inter-fraction errors occur frequently in interstitial HDR brachytherapy. If no action is taken, it will result in a significant risk of geometrical miss and overdose to the organs at risk. It is not recommended to use a single plan to deliver all the fractions. Imaging is recommended before each fraction and decision on re-planning must be taken.
引用
收藏
页码:143 / 151
页数:9
相关论文
共 50 条
  • [1] An Inter-Fraction Adaptive Strategy for High-Dose Rate Prostate Brachytherapy
    Pouliot, J.
    Foster, W.
    Cunha, A. J.
    Hsu, I-C
    [J]. MEDICAL PHYSICS, 2009, 36 (06)
  • [2] High-dose-rate interstitial brachytherapy for gynecologic malignancies
    Beriwal, Sushil
    Bhatnagar, Ajay
    Heron, Dwight E.
    Selvaraj, Raj
    Mogus, Robert
    Kim, Hayeon
    Gerszten, Kristina
    Kelley, Joseph
    Edwards, Robert P.
    [J]. BRACHYTHERAPY, 2006, 5 (04) : 218 - 222
  • [3] Interstitial high-dose-rate brachytherapy in eyelid cancer
    Mareco, Virginia
    Bujor, Laurentiu
    Abrunhosa-Branquinho, Andre N.
    Ferreira, Miguel Reis
    Ribeiro, Tiago
    Vasconcelos, Ana Luisa
    Ferreira, Cidalina Reis
    Jorge, Marilia
    [J]. BRACHYTHERAPY, 2015, 14 (04) : 554 - 564
  • [4] High-dose-rate interstitial brachytherapy for the treatment of penile carcinoma
    Petera, J
    Odrázka, K
    Zouhar, M
    Bedrosová, J
    Dolezel, M
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2004, 180 (02) : 123 - 125
  • [5] Perioperative interstitial high-dose-rate brachytherapy for keloids scar
    Barragan, Victoria Vera
    Garcia, Ana Isabel Alonso
    Garcia, Jose Fernandez
    Marin, Marta De Juan
    Pena Vivas, Johanna del Carmen
    Rijo, German Juan
    [J]. JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2022, 14 (01) : 29 - 34
  • [6] Inter-fractional variation of markers and applicators in single-implant high-dose-rate interstitial brachytherapy for gynecologic malignancies
    Ostyn, Mark
    Burke, Aidan M.
    Fields, Emma
    Todor, Dorin
    [J]. BRACHYTHERAPY, 2021, 20 (04) : 771 - 780
  • [7] Planning Margins and Inter-Fraction Motion for Inpatient Interstitial HDR Brachytherapy
    Campbell, W.
    Fisher, C.
    Miften, M.
    Schubert, L.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E575 - E576
  • [8] HIGH-DOSE-RATE BRACHYTHERAPY
    SPEISER, BL
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 : 116 - 116
  • [9] High-Dose-Rate Interstitial Brachytherapy as Monotherapy in Prostate Cancer Patients
    Kazberuk, D.
    Filipowski, T.
    Szmigiel-Trzcinska, A.
    Niksa, M.
    Hempel, D.
    Pancewicz-Janczuk, B.
    Nowik, W.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S448 - S448
  • [10] Interstitial high-dose-rate (HDR) brachytherapy in patients with sarcoma relapses
    Röddiger, S
    Sapoutzis, N
    Kurek, R
    Kolotas, C
    Filipowicz, I
    Vogt, HG
    Heyd, R
    Zamboglou, N
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2005, 181 : 65 - 65