A FERROFLUIDIC ACTUATOR FOR AN IMPLANTABLE ARTIFICIAL-HEART

被引:7
|
作者
MITAMURA, Y
WADA, T
SAKAI, K
机构
[1] HOKKAIDO UNIV,DEPT MED TECHNOL,SAPPORO,HOKKAIDO 060,JAPAN
[2] HOKKAIDO UNIV,SCH MED,SAPPORO,HOKKAIDO 060,JAPAN
关键词
FERROFLUIDIC ACTUATOR; ARTIFICIAL HEART;
D O I
10.1111/j.1525-1594.1992.tb00329.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The feasibility of using a ferrofluidic actuator for an artificial heart was studied. A ferrofluidic actuator directly drives magnetic fluids simply by applying a magnetic field to the ferrofluids and does not require a bearing. Magnetic fluid in a U-shaped glass cylinder was placed in an air gap of a solenoid. When a magnetic flux of 0.32 T was applied to the interface of the ferrofluid and air, the ferrofluid was displaced and a pressure of 7.58 kPa (57 mm Hg) was obtained. An array of three poles of solenoids was mounted near the U-shaped glass cylinder. The solenoids were sequentially activated and deactivated. A pressure of 9.98 kPa (75 mm Hg)/- 1.33 kPa (- 10 mm Hg) was obtained. Calculations indicate that a magnetic flux of 0.49 T is enough to obtain a pressure of 13.3 kPa (100 mm Hg). A ferrofluidic actuator is promising for use with an implantable artificial heart.
引用
收藏
页码:490 / 495
页数:6
相关论文
共 50 条