TECHNICAL CONSIDERATIONS IN CONTINUOUS JUGULAR VENOUS OXYGEN-SATURATION MEASUREMENT

被引:0
|
作者
DEARDEN, NM [1 ]
MIDGLEY, S [1 ]
机构
[1] WESTERN GEN HOSP,DEPT CLIN NEUROSCI,EDINBURGH EH4 2XU,MIDLOTHIAN,SCOTLAND
关键词
CEREBRAL ISCHEMIA; JUGULAR BULB OXYGEN SATURATION; REFLECTION OXIMETRY;
D O I
暂无
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Fibreoptic reflection oximetry allows continuous in-vivo estimation of jugular venous oxygen saturation. In combination with pulse oximetry the oxygen extraction ratio SaO2 - SjO2/SaO2 can be derived enabling identification of states of global luxury perfusion, normal coupling of global cerebral blood flow with global cerebral metabolism, global cerebral hypoperfusion and global cerebral ischaemia. Several technical difficulties may arise affecting the accuracy Of SjO2 recordings which must be recognised by the clinician before medical intervention is contemplated.
引用
收藏
页码:91 / 97
页数:7
相关论文