A RYDBERG-ATOM IONIZATION SOURCE FOR NEGATIVE-ION MASS-SPECTROMETRY

被引:0
|
作者
NAGATA, T
KONDOW, T
YAMAMOTO, Y
HAMAKAWA, S
AZAMI, K
机构
[1] UNIV TOKYO,FAC SCI,DEPT CHEM,BUNKYO KU,TOKYO 113,JAPAN
[2] NIPPON OIL CO LTD,CENT TECH RES LAB,NAKA KU,YOKOHAMA 231,JAPAN
关键词
RYDBERG-ATOM IONIZATION; NEGATIVE ION MASS SPECTROMETRY; 1-DECANETHIOL; 3-NITROBENZYL-ALCOHOL; O-NITROPHENYL OCTYL ETHER;
D O I
10.1016/0168-1176(93)87100-7
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
An ion source for negative ion mass spectrometry has been developed. The ionization method is based on the collisional electron transfer from a high-Rydberg atom (A**) to a target molecule which possesses a positive electron affinity. In this ionization process, negative ions are directly produced via resonance capture of the ''quasi-free'' Rydberg electrons: A** + M --> A+ + M-. In contrast to negative chemical ionization, the Rydberg-atom ionization proceeds at a low pressure of almost-equal-to 10(-3) Pa. The method was applied to several molecular systems, such as 1-decanethiol, 3-nitrobenzyl alcohol and o-nitrophenyl octyl ether.
引用
收藏
页码:217 / 223
页数:7
相关论文
共 50 条
  • [1] NEGATIVE-ION MASS-SPECTROMETRY
    DILLARD, JG
    CHEMICAL REVIEWS, 1973, 73 (06) : 589 - 643
  • [2] NEGATIVE-ION MASS-SPECTROMETRY
    HORNING, EC
    CARROLL, DI
    DZIDIC, I
    STILLWELL, RN
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1980, 179 (MAR): : 23 - ANAL
  • [3] NEGATIVE-ION MASS-SPECTROMETRY
    BRANDENBERGER, H
    FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1984, 317 (06): : 634 - 635
  • [4] NEGATIVE-ION MASS-SPECTROMETRY
    HORNING, EC
    CARROLL, DI
    DZIDIC, I
    STILLWELL, RN
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1979, (SEP): : 63 - 63
  • [5] NEGATIVE-ION MASS-SPECTROMETRY
    KARASEK, FW
    RESEARCH-DEVELOPMENT, 1974, 25 (04): : 34 - &
  • [6] PULSED POSITIVE NEGATIVE-ION CHEMICAL IONIZATION MASS-SPECTROMETRY
    HUNT, DF
    STAFFORD, GC
    CROW, FW
    RUSSELL, JW
    ANALYTICAL CHEMISTRY, 1976, 48 (14) : 2098 - 2105
  • [7] CALIBRATION SUBSTANCES FOR NEGATIVE-ION CHEMICAL IONIZATION MASS-SPECTROMETRY
    BRUINS, AP
    BIOMEDICAL MASS SPECTROMETRY, 1980, 7 (10): : 454 - 456
  • [9] NEGATIVE-ION CHEMICAL IONIZATION MASS-SPECTROMETRY OF SOME ISOCYANATES
    SUZUKI, O
    BRANDENBERGER, H
    FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1987, 326 (03): : 228 - 231
  • [10] NEGATIVE-ION DESORPTION MASS-SPECTROMETRY
    ANBAR, M
    STJOHN, GA
    JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1975, 97 (24) : 7195 - 7197